N4387A Linear Heat Series

Laser type Semi-conducting laser
Wavelength 1064 nm
Maximum average output power* 17 mW
Beam diameter 50 μm
Numeric Aperture 0.2
Laser class according to IEC 60825 1M
Maximum allowed average output power** 20 mW
Monitoring distances 1, 2, 3, 4, 8 km
Minimum sampling interval 0.5 m
Available measurement cycles 10 s to 30 s
Available settings for spatial resolution 1 m, 1.5 m, 3 m, 5 m, 8 m
Temperature resolution (standard deviation over distance) 30 s/ 1.5 m resolution +/- 2.5 °C
Temperature reproductability (standard deviation) 0.3 °C (1-4 km); 1 °C (8 km)
Temperature reproductability for whole operating temperature area (standard deviation) 2 °C
Absolute temperature standard deviation 2.5 °C
Available measuring modes Single-ended and dual-ended; including fiber break reconstruction
Operating voltage 10 V DC to 30 V DC
Current consumption at 20 °C Average 15 W, max < 40 W
Fiber types MM 50/125 μm MM 62.5/125 μm
Dynamic range 30 dB (loss in both directions)
Permissible operating temperature of sensor fiber -273 °C to +700 °C (possibly limited by sensor cable used)
Cable diameter
Sensor Cable Safety 4.0 mm
Sensor Cable Steel 3.8 mm
Cable weight
Sensor Cable Safety 17 kg/km
Sensor Cable Steel 25 kg/km
Permissible operating temperature
1 fiber channel -10 °C to +60 °C
2 fiber channels -5 °C to + 60 °C
Permissible storage temperature – 40 °C to +80 °C
Permissible relative humidity
1 fiber channel 0% to 95% non-condensing
2 fiber channels 15% to 85% non-condensing
Operation conditions for power supply (optional) 0 °C to 50 °C; non-condensing, only interior rooms
Measurements (H x W x D) 88 x 448 x 364 mm
Weight 9 kg
Optical connector E2000; 8° angled polish
Number of channels 1 or 2 channels
Computer interfaces USB, LAN
Relay board 4 inputs / 20 outputs*
Internal data storage capacity 150 measurements